Digital FPGA-based processing of pulses of gas-filled gamma-radiation detector for acoustic noise suppression
Abstract
Gas-filled proportional detectors of ionizing radiation often contain such elements as thin anode wire, which make them sensitive to external acoustic or vibrational impact. This study investigates such sensitivity for the case of the proportional γ-radiation detectors filled with high-purity xenon gas.
The detector demonstrates a dependence of its signal noise on the external acoustic wave frequency, the character of which most likely depends on the design of the detector, i.e. on such parameters as diameter and tension strength of the anode wire, etc. To suppress the negative impact of acoustic noise on the characteristics of the detector, the detector digital signal is processed in a Field-Programmable Gate Array board.
The authors investigate the algorithm of «time windows», which allows separating (filtering) low-frequency noise in the output signal of a gas-filled detector, which arises due to the influence of external sound, from the useful signal, i.e. the ionization pulses. This approach allows performing spectral measurements of γ-radiation with a significant acoustic noise background, when detectors with a conventional analog spectrometric channel cannot handle the task.
References
Vlasik K. F., Grachev V. M., Dmitrenko V. V. et al. [Gamma spectrometers on compressed xenon. ]. Instruments and Experimental Techniques, 1999, no. 5, pp. 114–122. (Rus)
Bolic M., Drndarevic V. Digital gamma-ray spectroscopy based on FPGA technology. Nuclear Instruments and Methods in Physics Research A, 2002, vol. 482, iss. 3, pp. 761–766. https://doi.org/10.1016/S0168-9002(01)01925-8
Becker E.M., Farsoni A.T. Wireless, low-cost, FPGA-based miniature gamma ray spectrometer. Nuclear Instruments and Methods in Physics Research A, 2014, vol. 761, pp. 99–104. https://doi.org/10.1016/j.nima.2014.05.096
Novikov A. S., Ulin S. E., Dmitrenko V. V. et al. New modification of xenon gamma-ray detector with high energy resolution. Optical Engineering, 2014, vol. 53, no. 2, 021108. https://doi.org/10.1117/1.OE.53.2.021108
Dmеtrenko V.V. et al. Vibrostability of high pressure xenon gamma-ray detectors. 1999 IEEE Nuclear Science Symposium. Conference Record. 1999 Nuclear Science Symposium and Medical Imaging Conference (Cat. No.99CH37019), 1999, vol. 2, pp. 665–669. https://doi.org/10.1109/NSSMIC.1999.845756
Seifert A., Milbrath B., Pitts W. K., Smith E. Implementation of a noise mitigation strategy for a high-pressure xenon detector. IEEE Nuclear Science Symposium Conference Record, 2005, pp. 1262–1266 https://doi.org/10.1109/NSSMIC.2005.1596549
Sokolov S., Pudov A., Rybka A. et al. Gas-filled gamma-radiation detector based on high-purity xenon. East European Journal of Physics, 2020, no. 4, pp. 28–34. https://doi.org/10.26565/2312-4334-2020-4-04
Manchanda R.K., Ye Z., Sood R.K. Ultra-high-pressure proportional counter. Part I: Argon. Nuclear Instruments and Methods in Physics Research A, 1990, vol. 292, pp. 373–385. https://doi.org/10.1016/0168-9002(90)90394-L
Copyright (c) 2022 Pudov A. O., Sokolov S. A., Abyzov A. S., Rybka A. V., Kutny V. E.

This work is licensed under a Creative Commons Attribution 4.0 International License.